Determination of Chromium in blood serum using Atomic Absorption Spectrophotometric (AAS)

Authors

  • Mohammed T. Khathi College of Science , Dhi- Qar University
  • Abeer Q. Abdulwahab College of Science , Dhi- Qar University
  • Zainab M. Kareem College of Science , Dhi- Qar University

Keywords:

Blood serum , Chromium ,Electrothermal and Tantalum.

Abstract

A new approach for the direct determination of chromium by Electrothermal Atomic Absorption spectrometry (ETAAS) is proposed, the method employs a 1.0 cm X 0.3 cm Tantalum metal coil placed in graphite tube. The experimental conditions for the determination of chromium via this method have been optimized. Pyrolysis and atomization curves are constructed using uncoated graphite tube without and with coil. The calibration curves were linear over the rang 10- 50 ng/ml of chromium concentration, with relative standard deviation values, at 0.013, 0.018 and 0.024µg/ml. the accuracy and precision of the method was evaluated by determining recoveries  percentage between (99.7-101) and standard deviation and  using different concentrations of standard solutions of  this method was applied successfully for the determination of  chromium in human blood serum.